End-Scan Technique Research

From ATTWiki
Revision as of 23:58, 10 February 2012 by Mcone (Talk | contribs)

Jump to: navigation, search

Vtube-laser logo.jpg End Scan Technique Research

Vtube-laser 1 73.jpg


End-Scan Techniques

Measure Setup

  • Set System Options, Measure 2 Tab, "Maximum Number of LLP Points in String" to 3
  • Set Measure Setup - LASER Tab: The LASER defaults to ON for CYLINDERS and END-SCANS.
  • Set Measure Setup - LASER Tab: Cylinder stripes = 3 to 5
  • Set Measure Setup - LASER Tab: End-scan stripes = 5
  • Set Measure Setup - LASER Tab: Split bend stripes = 5
  • Set Measure Setup = Scan Rates Tab: Set all rates to TIMER for all scan types.
  • Set Measure Setup = Scan Rates Tab: Set the Scan Rates TIMER is set to 0 for all TIMERS.

How To Use This End-Scan Technique

  • Align the scanner with a slight scissor angle across the end as is shown in the image on the right

  • Attempt to hold the laser motionless when collecting the stripes across the end to increase repeatability.

  • Consistently place the line across the same area of the tube end from measure to measure.

  • Press the green button until VTube-LASER is finished scanning all the end stripes.

EndWithScissorCloseup.jpg

EndWithScissor.jpg

Other Pages